6 adjusting the dot printing position (dot model) – Omega RD100B User Manual

Page 199

Advertising
background image

11-7

Maintenance

11

11.6 Adjusting the Dot Printing Position (Dot Model)

This section explains the details of adjusting the dot printing position on the chart paper.

It is recommended that the position be adjusted once a year to assure its recording

accuracy.

Note

Dot printing position is adjusted under standard operating conditions when the recorder is

shipped from the factory. Because the chart paper may expand or contract depending on the
environment in which the recorder is used, it is recommended that the position be readjusted

before use.

Procedure

1.

Warm up the recorder for at least 30 minutes.

2.

Check that the operating environment such as ambient temperature and

humidity is within the normal operating conditions (see section 12.6).

3.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

4.

Hold down the

and

keys simultaneously for 3 seconds to enter Basic

Setting mode.

5.

Press the

key or SHIFT +

key to select P_Adj and then press the

key.

6.

Press the

key to select Hysteresis (center of the chart paper), Zero (left

edge of the chart paper), or Full (right edge of the chart paper) and then press

the

key.

Adjust Hysteresis first followed by Zero and then Full.

Adjusting the Hysteresis

A line is drawn at the center of the chart paper. If the line appears as shown in Figure

a, increase the displayed setting. Then press the

key. If the line appears as shown

in Figure b, decrease the displayed setting. Then, press the

key.

To change the setting, press the

key to move the cursor and

key to change

the value.

Repeat this procedure until the line becomes straight.

A value change of 1 corresponds to 0.1 mm change in the dot printing position.

Up

Down

Down

Up

Figure a

Figure b

Adjusting Zero and Full

A line is drawn to the left edge for Zero and right edge for Full.

Increase or decrease the displayed setting so that the drawn line is aligned with the

scale line on the chart paper. Then, press the

key. The procedure of changing the

setting is the same as with the adjustment of hysteresis.

The line moves after the

key is pressed.

A value change of 1 corresponds to 0.1mm change in the dot printing position.

Decreasing the value moves the line to the left and increasing the value moves the

line to the right.

Advertising