6.12 Applying a Bias on the Measuring Input Signal

This section explains the details of setting a bias on the measuring input signal.

If you change the input range, set the bias again.

Procedure

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Press the

key or SHIFT +

key to select Bias and then press the

key.

3.

Set each item and press the

key.

Use the

key or SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.

If you press the ESC key in the middle of the operation, the settings entered up

to then are cancelled, and the display returns to a higher level menu.

4.

When the Setting complete screen appears, do either of the following:

Press the

key to set other channels.

If you are done, press the ESC key.

5.

Hold down the

MENU

key for 3 seconds to return to Operation mode.

Select whether to use the bias.

Set the bias value.

Set the channel range.

The new setting takes effect.

First channel

Last channel

ESC/?

(When bias is On)

Set=Bias

CH=01-01

Bias=Off

Bias= 0.000

01-01 Channel
Setting complete

Explanation

Channel Range

The only case when a bias can be set on multiple channels simultaneously is when the

channels are set to the same range type (for example channel 01 and 02 are set to 2 V

range). For channels on which scaling is set, the channels must be set to the same

range type, same span values, and same scaling values.

Bias

The range of bias that can be specified is ±10% of the span of the measurable range at the
specified range type. For example, the range is –0.4 V to 0.4 V for the range type 2 V.

For channels on which scaling is set, the range is ±10% of the scaling span.

Note

A bias cannot be set on channels set to ON/OFF input (DI).

<Related Topics>

Enabling the bias function: Section 7.15