Omega RD100B User Manual

Page 127

Advertising
background image

7-4

7.2

Changing the Integration Time of the A/D
Converter

This section explains the details of setting the integration time of the A/D converter.

Basic Setting mode cannot be entered when recording is in progress or when

computation is in progress on models with the computation function (/M1 option).

Procedure

Changing the Settings

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Hold down the

and

keys simultaneously for 3 seconds to enter Basic

Setting mode.

3.

Press the

key or SHIFT +

key to select A/D and then press the

key.

4.

Set each item and press the

key.

Use the

key or SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.

If you press the ESC key in the middle of the operation, the settings entered up

to then are cancelled, and the display returns to a higher level menu.

5.

When the Setting complete screen appears, do either of the following:

To correct the setting, press the

key.

If you are done, press the ESC key.

Select the integration time of the
A/D converter.

The new setting takes effect.

ESC/?

A/D Integrate
Setting complete

Basic=A/D

Integrate=Auto

Applying the Changes and Returning to Operation Mode

Press the

key or SHIFT +

key to select End and then press the

key. Press

the

key to select Store and then press the

key. The changes are applied, and

the screen returns to Operation mode. If you select Abort and press the

key, the

changes are discarded, and the screen returns to Operation mode. Press the ESC key

to return to the Basic= screen.

Explanation

Integrate

Selects the integration time of the A/D converter.

50Hz: Sets the integration time to 20 ms.

*

60Hz: Sets the integration time to 16.7 ms.

*

Auto:

Sets the integration time (20 ms or 16.7 ms) to a value synchronized with the

power supply frequency.

*

100ms (dot model): Sets the integration time to 100 ms. The scan interval is set to

2.5 s.

*

The scan interval is 125 ms (pen model) and 1 s (dot model) when set to 50Hz, 60Hz, or

Auto.