Setup Operations for Changing/Adding Functions (Basic Setting Mode)

7

Explanation

RJC

Sets the RJC mode.

Internal:

Uses the RJC function on the recorder.

External: Uses an external RJC function. If external is selected, set the compensation

voltage.

Volt

Sets the compensation voltage when using an external RJC function.

The compensation voltage can be set in the range of –20000 µV to 20000 µV.

7.4 Setting the RJC Function on Channels Set to TC Input