11.5 Adjusting the Pen Position (Pen Model)

This section explains the details of adjusting the pen position on the chart paper.

It is recommended that the pen be adjusted once a year to assure its recording accuracy.

Note

Pen position is adjusted under standard operating conditions when the recorder is shipped from

the factory. Because the chart paper may expand or contract depending on the environment in

which the recorder is used, it is recommended that the position be readjusted before use.

Procedure

1.

Warm up the recorder for at least 30 minutes.

2.

Check that the operating environment such as ambient temperature and

humidity is within the normal operating conditions (see section 12.6).

3.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

4.

Hold down the

and

keys simultaneously for 3 seconds to enter Basic

Setting mode.

5.

Press the

key or SHIFT+

key to select P_Adj and then press the

key.

6.

Select the Pen No.

7.

Press the

key to select Zero (left edge of the chart paper) or Full (right

edge of the chart paper) and then press the

key.

Adjust Zero first and then Full.

Increase or decrease the setting to align the pen position with the scale line on

the chart paper. Then, press the

key. To change the setting, press the

key to move the cursor and

key to change the value.

The pen moves after the

key is pressed.

A value change of 1 corresponds to 0.033 mm change in the pen position.

Decreasing the value moves the pen to the left and increasing the value moves

the pen to the right.

8.

Press the

key to set adjust other pens.

Press the ESC key to finish the adjustment. The screen returns to the

Basic=P_Adj

screen.

Select the pen No.

Select Zero or Full.

Change the value to adjust the pen
position.

The pen moves.

ESC/?

Basic=P_Adj

P_Adj=Zero

P_Adj
Setting complete

Pen No.=1

Zero=40

Applying the Changes and Returning to Operation Mode
Press the

key or SHIFT +

key to select End and then press the

key. Press

the

key to select Store and then press the

key. The changes are applied, and

the screen returns to Operation mode. If you select Abort and press the

key, the

changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.