7-15

Setup Operations for Changing/Adding Functions (Basic Setting Mode)

7

7.9

Setting the Bar Graph Display Mode

This section explains the details of setting the bar graph display mode.

Basic Setting mode cannot be entered when recording is in progress or when

computation is in progress on models with the computation function (/M1 option).

Procedure

Changing the Settings

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Hold down the

and

keys simultaneously for 3 seconds to enter Basic

Setting mode.

3.

Press the

key or SHIFT +

key to select Bar graph and then press the

key.

4.

Set each item and press the

key.

Use the

key or SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.

If you press the ESC key in the middle of the operation, the settings entered up

to then are cancelled, and the display returns to a higher level menu.

5.

When the Setting complete screen appears, do either of the following:

Press the

key to set other channels.

If you are done, press the ESC key.

Set the channel range.

Select the display mode of the bar graph.

The new setting takes effect.

ESC/?

First channel

Last channel

01-01 Channel
Setting complete

Basic=Bar graph

Graph=Normal

CH=01-01

Applying the Changes and Returning to Operation Mode

Press the

key or SHIFT +

key to select End and then press the

key. Press

the

key to select Store and then press the

key. The changes are applied, and

the screen returns to Operation mode. If you select Abort and press the

key, the

changes are discarded, and the screen returns to Operation mode. Press the ESC key

to return to the Basic= screen.

Explanation

Graph

Normal: Sets the base point of the bar graph to the left or right of the recording span

whichever is smaller in value.

Center: Sets the base point of the bar graph to the 50% position of the recording span.

Standard

Starting point of the bar graph

Starting point of the bar graph

Center-zero

Advertising