Input computation (standard function) – Omega RD100B User Manual

Page 204

Advertising
background image

12-2

Item

Specifications

Input type

Floating unbalanced input.
Isolation between channels (except, b terminal is shared for RTD input).

Burnout Detection of TC

Can be set for each channel.
Upscale/Downscale switchable.

Input Type

Operating Conditions

TC input

2 kΩ or less: normal, 10 MΩ or more: burnout, detection current: approx. 10 µA

1-5V input

0.2 V or less: burnout

RJC of TC input

Set whether to use the recorder RJC function or an external RJC function for each channel.
Set the compensation voltage when using an external RJC function.
Compensation voltage range: –20000 µV to 20000 µV.

A/D converter

Resolution

16 bits

Integration time

Pen model: Selectable from 20 ms (50 Hz), 16.7 ms (60 Hz), and Auto (auto switch between 20 ms

and 16.7 ms according to the power supply frequency).

Dot model: Selectable from 20 ms (50 Hz), 16.7 ms (60 Hz), 100 ms and Auto (auto switch

between 20 ms and 16.7 ms according to the power supply frequency).

Filter function (pen model) Filter ON/OFF switchable for each channel.

Time constant selectable from 2, 5, and 10 s.

Moving average function

Moving average ON/OFF switchable for each channel.

(dot model)

The number of samples of the moving average is selectable between 2 and 16.

Input Computation (Standard Function)

Item

Specifications

Delta computation

Calculates the difference between two channel inputs using the following equation.

Delta = (the input value of the channel set to delta computation) – (the input value of the
reference channel)
where (reference channel NO.) < (channel No. set to delta computation).

Computable input type: DC voltage, TC, and RTD. The channel set to delta computation and the
reference channel is set to the same range type.

Linear scaling

Scales the input value to a value with a unit suitable for the application.
Computable input type: DC voltage, TC, RTD, DI
Value after scaling

Mantissa: –20000 to 30000
Decimal position: Can be set arbitrarily
Unit: Can be set arbitrarily (up to 6 characters)

Displayable and printable range

–19999 to 31500

Computation on the 1-5V input

Linear scaling

Linearly scales the 1-5V signal.
Value after scaling: Same as linear scaling.
Displayable and printable range: Same as linear scaling.

Low-cut function

Sets values below 0% of the recording span to the leftmost value of the scale. The leftmost value of
scale must be less than the rightmost value of scale.

Square root computation

Takes the square root of the input value and linearly scales the result.
Computable input type: DC voltage

Linear scaling

Value after scaling: Same as linear scaling.
Displayable and printable range: Same as linear scaling.

Low-cut function

Sets values below the low-cut point to the leftmost value of scale. The leftmost value of scale must
be less than the rightmost value of scale.

Low-cut point range: 0.0 to 5.0% of the recording span (0.1% steps)

Bias

Adds a bias to the measuring input value.
Bias value range: ±10% of the span of the measurable range at the specified range type.

±10% of the scaling span if linear scaling is used.

12.1 Input Specifications