Arrangement of the measuring input terminals – Omega RD100B User Manual

Page 50

Advertising
background image

2-6

When using internal reference junction compensation on the thermocouple input,

take measures to stabilize the temperature at the input terminal.

• Always use the terminal cover.

• Do not use thick wires which may cause large heat dissipation (cross sectional area of

0.5 mm

2

or less recommended).

• Make sure that the ambient temperature remains reasonably stable. Large

temperature fluctuations can occur if a nearby fan turns ON or OFF.

Connecting the input wires in parallel with other devices can cause signal

degradation, affecting all connected devices.

If you need to make a parallel connection, then

• Turn the burnout detection function OFF.

• Ground the instruments to the same point.

• Do not turn ON or OFF another instrument during operation. This can have adverse

effects on the other instruments.

• RTDs cannot be wired in parallel.

Arrangement of the Measuring Input Terminals

A terminal cover is screwed in place on the measuring input terminal block on the rear

panel. A label indicating the terminal arrangement is affixed to the cover.

Measuring input terminal block

Terminal cover
attachment screws

Pen Model

+/A

–/B

b

+/A

–/B

b

Channel 1

Channel 2

Channel 3

Channel 4

Channel 1

Channel 3

Channel 2

Channel 4

+/A

–/B

b

Screw input terminal

Clamped input terminal

(/H2 option)

2.3 Input Signal Wiring