Item

Specifications

Reference junction compensation accuracy

Above 0°C with input terminal temperature balanced (60 minutes after power on)
Type R, S, B, W, WRe: ±1.0°C
Type K, J, E, T, N, L, U: ±0.5°C

Maximum input voltage

±10 VDC (continuous) for ranges of 200 mV or less, TC, RTD, and DI ranges
±60 VDC (continuous) for 2 VDC or higher ranges

Input resistance

Approximately 10 MΩ or more for ranges of 200 mV or less and TC
Approximately 1 MΩ for 2 VDC or higher ranges

Input source resistance

Volt, TC:

2 kΩ or less

RTD input:

10 Ω or less per wire (The resistance of all three wires must be equal).

Bias current

10 nA or less (except when burnout detection function is enabled)

Common mode rejection ratio 120 dB (50/60 Hz ± 0.1%, 500 Ω unbalanced, between the minus terminal and ground)
Normal mode rejection ratio

40 dB or more (50/60 Hz ± 0.1%)

Noise rejection

By the integrating A/D converter: The integration time is 16.7 ms, 20 ms, or 100 ms (dot model)
Low-pass filtering (pen model): Filter time constant is 2 s, 5 s, or 10 s
Moving average (dot model):

The number of samples is 2 to 16.

Effects of Operating Conditions

Item

Specifications

Ambient temperature

With temperature variation of 10°C:

Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:

Measurement fluctuation + ±0.2% of the recording span or less

* Excluding the error of reference junction compensation

Power supply fluctuation

With variation within 90 to 132 V and 180 to 264 VAC (50/60 Hz):

Measurement: ±1 digit or less
Recording:

±0.1% of the recording span or less

With variation of ±2 Hz from rated power frequency (power supply voltage 90 to 132, 180 to 264 VAC):

Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:

Same as the measurement fluctuation

Magnetic field

AC (50/60 Hz) and DC 400 A/m fields:

Measurement: ±(0.1% of rdg 10 digit) or less
Recording:

±0.5% of the recording span or less

Input source resistance

DC voltage range

With variation of +1 kΩ:

Ranges of 200 mV or less: Within ±10 µV
Ranges of 2V or more:

±0.1% of rdg

TC range

With variation of +1 kΩ: ±10 µV or less

RTD range

With variation of 10 Ω per wire (resistance of all three wires must be equal):

Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:

Measurement fluctuation + ±0.1% of the recording span or less

With maximum difference of 40 mΩ between wires: Approx. 0.1°C (for Pt100)

Mounting position

With a backward tilt of 30° or less

Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:

Indication fluctuation + ±0.1% of the recording span or less

Vibration

When a sinusoidal vibration along all three axis at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s

2

is applied for 2 hours

Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:

Indication fluctuation + ±0.1% of the recording span or less

Other Specifications

Item

Specifications

Clock

With a calendar function

Accuracy of clock

±100 ppm, excluding a delay (of 1 second, maximum) caused each time the power is turned on.

DST

The date/time for switching between standard time and DST can be specified.

Memory backup

A built-in lithium battery backs up the setup parameters and runs the clock
(battery life: approximately ten years at room temperature).

Key lock function

Password for releasing the key lock can be set.

Key lock items

RCD key, MENU key, DISP MENU key, FEED key, and function under the FUNC key (Alarm ACK, computation start/
stop, print, message, printout buffer clear, periodic printout reset, and pen or ribbon cassette exchange)

Internal light

Lights up the recording area of the chart paper. Uses a dedicated LED.

Noise

Machine Noise Information Ordinance 3.GSGV, Jan 18, 1991
Maximum noise level: 60 dB (A) or less (complies with ISO7779)

12.6 General Specifications