Programming Function Blocks and Features

Video Recorder – User Manual

132

Sterilization

The output CVn OV is a “kill time”, F

0

, which quantifies the destruction of microorganisms during a

sterilization process. F

0

represents the number of minutes after which the microorganism population is

decreased by one decimal or log unit. The F

0

formula is:

F

0

=

∑ [dt x (10

(T-Tref)/Z

)]

where:

dt = time between F

0

measurements, in minutes. Dt is equal to the scan cycle of the instrument in

minutes.

T = measured temperature input, in °C

T

ref

= reference temperature (121°C typical)

Z = product-dependent temperature constant

This function also compares the calculated F

0

value to the desired F

0

value, F

d

(Preset). When F

0

equals or exceeds F

d

, the function activates its discrete output CVn OS; this output remains ON(1) until

the F

0

calculation is reset via the Reset Input.

The output CVn A2 is the Time Remaining (Tr) at the current temperature to achieve F

d

and is

calculated as follows:

(F

d

– F

0

)

Tr =

_________

10

(T-Tref)/Z

The F

0

calculation will be reset and the F

0

value will remain at zero while the Reset Input is ON(1). The

calculation will resume when the Reset Input goes OFF(0) again.

Table 4-40 describes the Sterilization prompts. See Table 4-2 for additional prompts.

Table 4-40 F

0

Sterilization Prompts

Prompt

Range/Selections

Definition

OUT LOW LIMIT

OUT HIGH LIMIT

OFF or number

Displayed output value will be clamped and will flash
when these limits are exceeded.

RESET

OFF, discrete parameter, 0,
1

Reset input

TEMP INPUT

OFF, analog parameter,
number

Measured temperature in degrees C

REFERENCE TEMP

OFF, analog parameter,
number

Reference temperature

TEMP CONSTANT

OFF, analog parameter,
number

Temperature constant Z

DESIRED F0

OFF, analog parameter,
number

Desired F

0

(F

d

)