4 unit for displayed temperature, Unit for displayed temperature -15, Operation – Yokogawa EJX440A User Manual

Page 25

Advertising
background image

<3. Operation>

3-15

IM 01C25T01-01E

Note that following symbols are not available:

# % & < > . * : + -

The integral indicator shows “-- -- -- -- -- --” when

these are entered.
Engr LRV and Engr URV are used to set the lower

and upper range values for the engineering unit

display. When the insrument is shipped, these are

set as specified in the order.

F0323.ai

Call up the

Engr LRV Display.

Set

–50, and press ENTER (F4).

EJX:YOKOGAWA

Engr LRV

0.0

-50

DEL

DEL

ESC

ENTER

EJX:YOKOGAWA

Engr disp range

1 Engr LRV -50

2 Engr URV 100

3 Engr exp ×1

4 Engr Unit M

5 Engr point 1

HELP

SEND

HOME ENTER

Press to select engr disp URV.

EJX:YOKOGAWA

Engr URV

100.0

50

DEL

DEL

ESC

ENTER

Set

50, and press ENTER (F4).

HELP

(ENTER)

‘– 5 0’

(ENTER)

‘5 0’

(SEND)

1. Device setup
4. Detailed setup
4. Display condition
4. Engr disp range
1. Engr LRV
and 2. Engr URV

Example: Set lower range value (LRV) to

–50 and upper

range value (URV) to

50.

1

2

EJX:YOKOGAWA

Engr disp range

1 Engr LRV -50

2 Engr URV 50

3 Engr exp ×1

4 Engr Unit M

5 Engr point 1

HELP

SEND

HOME ENTER

4

3

Press

SEND (F2) to send the data

to the transmitter, then check to

confirm that disappears.

(SEND)

SEND

3.5.4 Unit for Displayed Temperature

When the instrument is shipped, the temperature

units are set to

C (Centigrade). Follow the

procedure below to change this setting.
When this parameter is set, it also changes the

temperature unit for

Snsr temp at Process

variables and Amp temp at Temp sensor.

F0325.ai

(ENTER)

EJX:YOKOGAWA

Temp Unit

degC

degC

degF

Kelvin

HELP

SEND

ESC

ENTER

Select

degF (Fahrenheit), and

Press

ENTER (F4).

1. Device setup
4. Detailed setup
1. Sensors
3. Temp sensor
3. Temp Unit

Example: Change the unit for the temperature display from

degC to degF.

EJX:YOKOGAWA

Temp sensor

1 Snsr temp 23 degC

2 Amp temp 23 degC

3 Temp Unit degF

HELP

SEND

HOME ENTER

2

1

Press

SEND (F2) to send the data

to the transmitter, then check to

confirm that disappears.

(SEND)

SEND

Advertising
This manual is related to the following products: