<4. Diagnostics>

4-1

IM 01C25T01-01E

4. Diagnostics

4.1 Self-Diagnostics

4.1.1 Identify Problems by Using the

Communicator

The HART communicator can be used to run self-

diagnostics on a transmitter and check for incorrect

data settings.
The

Self test and Status commands are available

for self-diagnostics. When

Self test is run, the

integral indicator shows an error code and alarm

message if the transmitter detects any illegal

parameter settings and functional faults. See

table 4.3.1 Alarm Message Summary for probable

causes and countermeasures. If the specific

diagnostic item is known for the check, you can

directly call up the item by using the

Status

command. See table 4.3.1 to determine the status

group.
The HART communicator diagnoses every

command you make. When a faulty command

or keypad input is performed, an error message

appears. See table 4.3.2 HART Commnicator Error

Messages for the details.

Diagnostic by “self test”

1

F0401.ai

Call up

Test device, and select

Self test.

EJX:

Test device

1 Status

2 Self test

HELP

SAVE

HOME

OK

EJX:

Self test OK

DEL

SET

ABORT

OK

2

If no error is detected,

Self test

OK is displayed.

If there is an error, an error

message appears.

1. Device setup
2. Diag/Service
1. Test device
2. Self test

Diagnostic by “status”

1

F0402.ai

Call up

Status, and select Status

group 6.

EJX:

Status

2 Status group 2

3 Status group 3

4 Status group 4

5 Status group 5

6 Status group 6

HELP

SAVE

HOME

OK

EJX:

Status group 6

Illegal P LRV On

Illegal P URV Off

Illegal P SPAN Off

P SPAN trim err Off

P ZERO trim err Off

DEL

SET

ABORT EXIT

2

If there is no error, the result of

diagnostics is indicated as

Off.

If

On is indicated, a

countermeasure for that error is

necessary.

1. Device setup
2. Diag/Service
1. Test device
1. Status
6. Status group 6

4.1.2 Checking with Integral Indicator

NOTE

If an error is detected by running self-diagnostics,

an error number is displayed on the integral

indicator. If there is more than one error, the error

number changes at three-second intervals. See

table 4.3.1 regarding the alarm codes.

F0403.ai

Figure 4.1.2 Integral Indicator