<5. Parameter Summary>

5-5

IM 01C25T01-01E

Function

Label

Item

Contents

Default value

Handling

Fast key

sequences

Static

pressure

setup

A/G Select

Gauge/abs select for static pressure

Gauge or Absolute

Absolute

WD

1, 4, 1, 2, 7, 4, 1

Atm. Pres value

Conversion coefficient

101.3 kPa

WD

1, 4, 1, 2, 7, 4, 2

SP H/L Select

H/L select for static pressure

High or Low

High

WD

1, 4, 1, 2, 7, 5

SP Apply values

Rerange for static pressure

“0%, 100%, or Exit”

MD

1, 4, 1, 2, 7, 1, 2

SP Damp

Damping time constant for SP

0.00 to 100.00 sec

2.00 sec

WD

1, 4, 1, 2, 7, 3

SP LRV

Lower range value for static pressure Within measurement range

0.0 MPa

WD

1, 4, 1, 2, 7, 1,

1, 1

SP LSL

Lower sensor limit for static pressure

RD

1, 4, 1, 2, 7, 1,

1, 4

SP Min Span

Minimum span for static pressure

RD

1, 4, 1, 2, 7, 1,

1, 6

SP URV

Upper range value for static

pressure*

2

Within measurement range

WD

1, 4, 1, 2, 7, 1,

1, 2

SP USL

Upper sensor limit for static pressure

RD

1, 4, 1, 2, 7, 1,

1, 5

Status

Status group 1

Device status information for

hardware

R

1, 2, 1, 1, 1

Status group 2

Device status information for

hardware

R

1, 2, 1, 1, 2

Status group 3

Device status information for process

R

1, 2, 1, 1, 3

Status group 4

Device status information for process

R

1, 2, 1, 1, 4

Status group 5

Device status information for process

R

1, 2, 1, 1, 5

Status group 6

Device status information for settings

R

1, 2, 1, 1, 6

Status group 7

Device status information for settings

R

1, 2, 1, 1, 7

Status group 8

Device status information for settings

R

1, 2, 1, 1, 8

Status group 9

Device status information for settings

R

1, 2, 1, 1, 9

Tag

Tag

Tag number

8 alphanumerics

As specified

W

1, 3, 1

Temperature

compensation

T.Z. Cmp mode

Temperature compensation mode

Off or On

Off

W

1, 4, 2, ↓, 1

Temp Zero

Zero shift compensation

–99.999 to 99.999%/degC

0.000%/degC

W

1, 4, 2, ↓, 2

Temperature

sensor

Amp temp

Amplifier temperature

R

1, 4, 1, 3, 2

Snsr temp

Capsule temperature

R

1, 4, 1, 3, 1

Temp Unit

Temperature setting unit

degC, degF, or Kelvin

degC

W

1, 4, 1, 3, 3

Test key

Test Key

Special maintenance parameter

M

1, 4, 6

Transfer

function

Xfer fnctn

Output mode

Linear or Sq root

As specified or

Linear

W

1, 3, 5

Unit

Unit

Measurement range unit

See section 3.4.2

As specified or

kPa

W

1, 3, 2

SP Unit

Static pressure unit

See section 3.5.5

MPa

WD

1, 4, 1, 2, 3

Write protect

menu

Enable wrt 10min

Write protect release

8 alphanumerics

M

Hot key, 2, 2

New password

User set password for write protect

8 alphanumerics

M

Hot key, 2, 3

Software seal

Software seal

Keep or Break

Keep

R

Hot key, 2, 4

Write protect

Write protect indicator

Yes or No

No

R

Hot key, 2, 1

*1:

Handling:

R=Read only, W=Read & Write, M=Method, A=Applicable for option code AL, D=Applicable for differential pressure

transmitters. Do not change these parameters for pressure transmitters.

*2:

The default value shows MWP (Maximum working pressure) of the capsule.

Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.