<5. Parameter Summary>

5-2

IM 01C25T01-01E

Function

Label

Item

Contents

Default value

Handling

Fast key

sequences

Diag Lim

Lim fDPmax

Upper limit for Ratio fDP

As specified

model

W, G

1, 2, 4, 3, 2, 3, 1

Lim fDPmin

Lower limit for Ratio fDP

As specified

model

W, G

1, 2, 4, 3, 2, 3, 2

Lim fSPlmax

Upper limit for Ratio fSPl

As specified

model

W, G

1, 2, 4, 3, 2, 3, 3

Lim fSPlmin

Lower limit for Ratio fSPl

As specified

model

W, G

1, 2, 4, 3, 2, 3, 4

Lim fSPhmax

Upper limit for Ratio fSPh

As specified

model

W, G

1, 2, 4, 3, 2, 3, 5

Lim fSPhmin

Lower limit for Ratio fSPh

As specified

model

W, G

1, 2, 4, 3, 2, 3, 6

Lim BlkFmax

Upper limit for BlkF

As specified

model

W, G

1, 2, 4, 3, 2, 3, 7

Lim BlkFmin

Lower limit for BlkF

As specified

model

W, G

1, 2, 4, 3, 2, 3, 8

Lim DPAvgmax

Upper limit for DPAvg

As specified

model

W, G

1, 2, 4, 3, 2, 3, 9

Lim DPAvgmin

Lower limit for DPAvg

As specified

model

W, G

1, 2, 4, 3, 2, 3, a

Diag Mode

Diag Mode

ILBD operation mode

Stop, Calculation, or

Reference

W, G

1, 2, 4, 3, 1, 1

Diag Option

Diag Option

Alarm masking

W, G

1, 2, 4, 2

Diag Output

Diag Out Option

Output mode of 4-20mA when

an advanced diagnostic alarm is

generated

Off, Burnout, or Fall back

Off

W, G

1, 2, 4, 5, 1

Diag Fixed Out Val 4-20 mA output when an advanced

diagnostic alarm is generated

3.6000 to 21.6000 mA

21.6 mA

W, G

1, 2, 4, 5, 2

Diag Period Diag Period

Sampling period per one Diag count

180 sec

W, G

1, 2, 4, 3, 2, 2

Diag

Reference

Diag Description

Memo field

32 alphanumerics

W, G

1, 2, 4, 3, 2, 4, 1

Ref fDP

Reference value of fDP

W, G

1, 2, 4, 3, 2, 4, 2

Ref fDP Status

Status of Reference fDP

R, G

1, 2, 4, 3, 1, 4, 2

Ref fSPl

Reference value of fSPl

W, G

1, 2, 4, 3, 2, 4, 3

Ref fSPl Status

Status of Reference fSPl

R, G

1, 2, 4, 3, 1, 4, 4

Ref fSPh

Reference value of fSPh

W, G

1, 2, 4, 3, 2, 4, 4

Ref fSPh Status

Status of Reference fSPh

R, G

1, 2, 4, 3, 1, 4, 6

Ref BlkF

Reference value of BlkF

W, G

1, 2, 4, 3, 2, 4, 5

Ref BlkF Status

Status of Reference BlkF

R, G

1, 2, 4, 3, 1, 4, 8

Ref DPAvg

Reference value of DPAvg

W, G

1, 2, 4, 3, 2, 4, 6

Ref DPAvg Status

Status of Reference DPAvg

R, G

1, 2, 4, 3, 1, 4, ↓

Diag Ref Lim Ref Lim fDPmin

Lower limit of fDP

7.0E-10

W, G

1, 2, 4, 3, 2, 6, 1

Ref Lim fSPmin

Lower limit of fSPl and fSPh

1.0E-10

W, G

1, 2, 4, 3, 2, 6, 2

Ref Lim BlkFmax

Upper limit of BlkF

0.5

W, G

1, 2, 4, 3, 2, 6, 3

Diag Supp

Count

Diag Supp Count

Detection count to generate an alarm

W, G

1, 2, 4, 3, 2, 5

Diag

Variables

Ratio fDP

SQRT (fDP/Ref fDP).

R, G

1, 2, 4, 3, 1, 3, 2

Ratio fDP Status

Status of Ratio fDP

R, G

1, 2, 4, 3, 1, 3, 1

Ratio fSPl

SQRT (fSPl/Ref fSPl).

R, G

1, 2, 4, 3, 1, 3, 4

Ratio fSPl Status

Status of Ratio fSPl

R, G

1, 2, 4, 3, 1, 3, 3

Ratio fSPh

SQRT (fSPl/Ref fSPl).

R, G

1, 2, 4, 3, 1, 3, 6

Ratio fSPh Status

Status of Ratio fSPh

R, G

1, 2, 4, 3, 1, 3, 5

BlkF

Blockage degree characterized in

comparison of high-pressure side

and low-pressure side pressure

fluctuation values

R, G

1, 2, 4, 3, 1, 3, 8

BlkF Status

Status of BlkF

R, G

1, 2, 4, 3, 1, 3, 7

DPAvg

Ratio of the average of differential

pressure/pressure to the maximum

span of an EJX transmitter.

R, G

1, 2, 4, 3, 1, 3, ↓

DPAvg Status

Status of DPAvg

R, G

1, 2, 4, 3, 1, 3, 9

CRatio fDP

Compnesated fDP

R, G

1, 2, 4, 3, 1, 3, ↓

CRatio fDP

Status of Cratio fDP

R, G

1, 2, 4, 3, 1, 3, ↓

NRatio fDP

Non-compnesated fDP

R, G

1, 2, 4, 3, 1, 3, ↓

NRatio fDP Status

Status of Nratio fDP

R, G

1, 2, 4, 3, 1, 3, ↓

*1:

Handling:

R=Read only, W=Read & Write, M=Method, A=Applicable for option code AL, G=Applicable for option code DG6,

D=Applicable for differential pressure transmitters. Do not change these parameters for pressure transmitters.

*2:

The default value shows MWP (Maximum working pressure) of the capsule.

Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.