8 capability test of blockage detection operation, Important – Yokogawa EJX440A User Manual

Page 52

Advertising
background image

<4. Diagnostics>

4-12

IM 01C25T01-01E

Start of Sampling

The sampling of reference value is carried out for

180 sec, which is the default value set to

Diag

Period.
1) Confirm that the sampling period (

Diag Period)

is set to 180 sec.

2) Set “Reference” to

Diag Mode.

The sampling starts soon after the setting.

IMPORTANT

• For the each parameter, the one value is

given. If Reference is set to

Diag Mode

again, the value is updated and overwritten.

• If the power supply is shut down during the

sampling,

Diag Mode becomes “Stop”. Set

“Reference” to

Diag Mode in order to carry

out the sampling again.

End of Sampling

After about 180 sec, the sampling automatically

finishes. The “Reference” setting of

Diag Mode

moves automatically to “Calculation”.
Confirm that the setting of

Diag Mode moves to

“Calculation”.

Reference Values

Confirm the latest values are obtained into the

following parameters.

• Ref fDP

• Ref fSPl

• Ref fSPh

• Ref BlkF

• Ref DPAvg

Invalid Ref F, SPH, SPL, or DP

When the enough reference fluctuation value is

not obtained, an alarm of invalid reference value

for each parameter is generated and also the ILBD

operation is not carried out.
Confirm the alarm of

Invalid Ref F, SPH, SPL, or

DP is not displayed in Diag Error.
If an alarm of

Invalid Ref F, SPH, SPL, or DP is

generated, consider the process condition or obtain

the reference fluctuation values again.

NOTE

Even if an alarm of

Invalid Ref F, SPH, SPL, or

DP is generated, “Calculation” in Diag Mode is

kept.

4.2.2.8 Capability Test of Blockage

Detection Operation

Before performing the ILBD operation, check the

capability of the blockage detection operation.

The simulation test is performed by closing motion

of a three-valve manifold or stop valve. When

simulated blockage occurs, confirm that an alarm is

generated.

NOTE

• The fluctuation amplitude of atmospheric

pressure is nearly zero with pressure or

level measurement. In such case, simulate

the blockage detection by closing the valve

where the fluctuation existed.

• Use Device description (DD) file for

parameter setting.

Simulation of High-pressure Side

Blockage

1) Close the high-pressure-side valve.

2) Confirm the value of

Pres is stable. If not,

open the valve a little.

3) Set “Calculation” to

Diag Mode so as to start

blockage detection operation.

4) Check that an alarm of “H Side Blocking” is

generated after the time that consists of

Diag

Period and Diag Supp Count passed.

5) Check also the operation of the analog alert if

an analog alert is set.

6) Open the valve completely and check that

there are no alarms.

Simulation of Low-pressure Side

Blockage

1) Close the low-pressure-side valve.

2) Confirm the value of

Pres is stable. If not,

open the valve a little.

3) Set “Calculation” to

Diag Mode so as to start

blockage detection operation.

4) Check that an alarm of “L Side Blocking” is

generated after the time that consists of

Diag

Period and Diag Supp Count passed.