Yokogawa EJX440A User Manual

Page 28

Advertising
background image

<3. Operation>

3-18

IM 01C25T01-01E

1

1. Device setup
2. Diag/Service
3. Calibration
3. Pres sensor trim
2. Pres Trim

EJX:YOKOGAWA

Select trim mode

1 Off

2 Auto, Lower Pt

3 Auto, Upper Pt

4 Manual, Lower Pt

5 Manual, Upper Pt

ABORT ENTER

F0330.ai

Select the

Auto, Lower Pt, and

press

ENTER (F4).

(ENTER)

2

EJX:YOKOGAWA

Pres for trim 1383.0

Auto, Lower Pt
0.000000

1350

ABORT

DEL

ENTER

Enter the value of the actual level

(1350 mmH

2

O), and press

ENTER (F4).

(ENTER)

‘1 3 5 0’

c. Using External Zero-adjustment Screw

This method permits zero adjustment without

the HART communicator. Use a slotted

screwdriver to turn the zero-adjustment screw.

See the hardware manual for details.
Note that the parameter of

Ext SW must be

Enabled to perform this adjustment. See

section 3.5.11 for the setting procedure.

(2) Full Sensor Trim—Auto Trim and Manual

Trim

Full sensor trim is carried out with a series of the

procedure of

Auto, Lower Pt and Auto, Upper

Pt. Also, you can manually perform the trimming

procedure in

Manual, Lower Pt and Manual,

Upper Pt.
The full sensor trim is a two-point adjustment,

and the lower point adjustment should always be

performed before the upper point adjustment in

order to maintain the pitch between the zero and

100% points within the calibration range.
In the manual method, the reference pressure

should also be applied to the transmitter at both

lower and upper point of trim ends. Without the

reference pressure,

P LTD and P UTD may not

represent the correct value of adjustment point for

each.

a. Auto Sensor Trim

Apply a standard pressure of 1000

mmH

2

O to the transmitter. After

obtaining a stable pressure, press

ENTER

(F4).

Example: For the range of 1000 to 3000 mmH

2

O

F0331.ai

1

EJX:YOKOGAWA

Select trim mode

1 Off

2 Auto, Lower Pt

3 Auto, Upper Pt

4 Manual, Lower Pt

5 Manual, Upper Pt

ABORT ENTER

Select

Auto , Lower Pt, and

press

ENTER (F4).

(ENTER)

2

EJX:YOKOGAWA

Pres for trim 994.0

Auto, Lower Pt
1000.000000

1000

ABORT

DEL

ENTER

(ENTER)

‘1000’

Apply a standard pressure of

3000 mmH

2

O to the transmitter.

After obtaining a stable pressure,

press

ENTER

(F4).

3

EJX:YOKOGAWA

Select trim mode

1 Off

2 Auto, Lower Pt

3 Auto, Upper Pt

4 Manual, Lower Pt

5 Manual, Upper Pt

ABORT ENTER

Select

Auto, Upper Pt, and

press

ENTER (F4).

(ENTER)

4

EJX:YOKOGAWA

Pres for trim 3015.0

Auto, Upper Pt
3000.000000

3000

ABORT

DEL

ENTER

(ENTER)

‘3000’

1. Device setup
2. Diag/Service
3. Calibration
3. Pres sensor trim
2. Pres Trim

Ч2

Advertising
This manual is related to the following products: