11 external switch mode, External switch mode -23 – Yokogawa EJX440A User Manual

Page 33

Advertising
background image

<3. Operation>

3-23

IM 01C25T01-01E

NOTE

1. When the polling option is set as

Never Poll

or

Poll Using Tag, the online menus cannot

be called up and displayed. Be sure to select

a polling option such as

Ask Before Polling.

2. When the same polling address is set for

two or more transmitters in multidrop mode,

communication with these transmitters is

disabled.

1

1

1

F0337.ai

(1) The HART communicator

searches for a transmitter that

is set in multidrop mode when

it is turned on.

When the HART

communicator is connected to

the transmitter, the polling

address and the tag will be

displayed (display ).

(2) Select the desired transmitter.

After that, normal

communication to the selected

transmitter is possible.

However, the communication

speed will be slow (display ).

(3) To communicate with another

transmitter, turn off the power

once and then turn on it again,

or call up display and select

Online.

(4) Display will appear. Select

the desired transmitter.

HART Communicator

Online

1 EJX110A-1

2 EJX110A-2

3 EJX110A-3

HELP

SET

HOME

OK

EJX:EJX110A-1:

Online

1 Device setup

2 Pres 0.0 mmH2O

3 AO 4.000 mA

4 LRV 0.0 mmH2O

5 URV 3500.0 mmH2O

DEL

SET

ABORT

OK

2

2

HART Communicator

1 Offline

2 Online

3 Frequency Device

4 Utility

HELP

DEL

ABORT

OK

3

3

Example: Communication when set in multidrop mode

To release multidrop mode, follow the procedure

below.
1. Call up the

Poll addr display and set the

address to 0.

2. Call up the

Polling display and set Ask Before

Polling.

3.5.11 External Switch Mode

Follow the procedure below to enable or inhibit zero

point adjustment by means of the zero-adjustment

screw on the transmitter.
This is set to

Enabled when the instrument is

shipped.

F0338.ai

(ENTER)

EJX:

Ext SW

Enabled

Disabled

Enabled

HELP

SEND

ESC

ENTER

Select

Disabled and press

ENTER (F4).

1. Device setup
4. Detailed setup
5. Device information
1. Field device info
9. Ext SW

Example: Set the mode to inhibit zero adjustment by

means of the external zero-adjustment screw.

EJX:

Field device info

5 Model 1

6 Model 2

7 Model 3

8 Write Protect No

9 Ext SW Disabled

HELP

SEND

HOME ENTER

Press

SEND (F2) to send the data

to the transmitter, then check to

confirm that disappears.

(SEND)

SEND

2

1