B. manual sensor trim – Yokogawa EJX440A User Manual

Page 29

Advertising
background image

<3. Operation>

3-19

IM 01C25T01-01E

b. Manual Sensor Trim

Suppose that a standard pressure of 1000 mmH

2

O is

applied and the value of the Pres for Trim in is 994.0.

Correct for this output error of 6 mmH

2

O by adding 6

mmH

2

O to

P LTD.

−4.0+6.0=+2.0

Enter the correction value of 2.

Then press

ENTER (F4).

Suppose that a standard pressure of 3000 mmH

2

O is

applied and the value of the Pres for Trim in is 3015.0.

Firstly, obtain the slope error for the span as follows;

Then correct for this slope error of −10 by adding −10 to

P UTD.

−3.0 + (−10.0) = −13.0

F0332.ai

1

EJX:YOKOGAWA

Select trim mode

1 Off

2 Auto, Lower Pt

3 Auto, Upper Pt

4 Manual, Lower Pt

5 Manual, Upper Pt

ABORT ENTER

Select

Manual, Lower Pt, and

press

ENTER (F4).

(ENTER)

2

2

EJX:YOKOGAWA

Pres for trim 994.0

Manual, Lower Pt

-4.000000

2

ABORT

DEL

ENTER

(ENTER)

‘2’

Enter the correction value of −13.

Then press

ENTER (F4).

3

EJX:YOKOGAWA

Select trim mode

1 Off

2 Auto, Lower Pt

3 Auto, Upper Pt

4 Manual, Lower Pt

5 Manual, Upper Pt

ABORT ENTER

Select

Manual, Upper Pt, and

press

ENTER (F4).

(ENTER)

4

4

EJX:YOKOGAWA

Pres for trim 3015.0

Manual, Upper Pt

-

3.000000

-

13

ABORT

DEL

ENTER

(ENTER)

13’

1. Device setup
2. Diag/Service
3. Calibration
3. Pres sensor trim
2. Pres Trim

Example: For the range of 1000 to 3000 mmH

2

O

P LTD = −4.0 mmH

2

O

P UTD = −3.0 mmH

2

O

Applied Pressure Value – Value of Pres for Trim

Applied Pressure Value

Slope Error =

×(URV–LRV)

3000 − 3015

3000

Ч (3000 − 1000)

=

= −10

Ч3

Ч4

(3) Sensor Trim for Static Pressure
For the EJX differential transmitters, zeroing and

full sensor trim of the static pressure is performed in

the same way as with the primary process variable

(PV). Note that the static pressure sensor trim

should be done only after trimming the PV.

(4) Reset Trim Adjustment to Factory Setting
The

Clear P snsr trim and Clear SP snsr trim

commands can reset the trim adjustment to the

initial calibrated values that were set. The amount

of the adjustment by the external zero-adjustment

screw is returned to the initial setting as well.