<1. Introduction>

1-4

IM 01C25T01-01E

1.4 Matching of Communicator

DD and Instrument DD

CAUTION

Before using the 275 HART Communicator,

make that the device description(DD) installed in

the communicator matches that of the instrument

that is being set up. To check the DD of the

instrument and the HART communicator, follow

the steps below. If the correct DD is not installed

in the communicator, you must upgrade the DD

at an authorized facility. For communication tools

other than the 275 HART Communicator, contact

the vendor for upgrade information.
1. Checking the DD of the instrument

1) Connect the communicator to the instrument

that is being set up.

2) Call

Device setup and press [→].

3) Call

Review and press [→].

4) Press [NEXT] or [PREV] to display

Fld dev

rev to show the DD of the instrument.

F0101.ai

EJX:

Review

Fld dev rev

1

HELP

PREV

NEXT

EXIT

The instrument DD

version is 1.

[Example]

2. Checking the DD of the 275 HART

Communicator

1) Turn on only the communicator.
2) Call

Utility from the main menu and press

[→].

3) Call

Simulation and press [→].

4) Select

YOKOGAWA from the list of

manufacturers by pressing [ ↓ ] and press

[→].

5) Select the model name of the instrument(i.e.

EJX) by pressing [ ↓ ] and press [→] to show

the DD of the communicator.

F0102.ai

HART Communicator

Fld dev rev

1 Dev v1, DD v1

[Example]

The communicator DD

supports Version 1.