<4. Diagnostics>

4-19

IM 01C25T01-01E

4.3 Alarms and Countermeasures

Table 4.3.1 Alarm Message Summary

Integral

indicator

HART communicator

display

Cause

Output operation

during error

Countermeasure

Status

group

AL. 01

CAP. ERR

P sensor error

Sensor problem.

Outputs the signal

(High or Low) set

with burnout direction

switch. [status output:

undefined]

Replace capsule when

error keep appearing

error even after restart.

1

CT sensor error

Capsule temperature

sensor problem.

Replace capsule.

Cap EEPROM error

Capsule EEPROM

problem.

AL. 02

AMP. ERR

AT sensor error

Amplifier temperature

sensor problem.

Replace amplifier.

Amp EEPROM error Amplifier EEPROM

problem.

CPU board error

Amplifier problem.

2

No device ID

No device ID is found.

Continues to operate

and output.

AL. 10

PRESS

P outside limit

Input is outside

measurement range

limit of capsule.

Outputs AO upper limit

or AO lower limit.

Check input or

replace capsule when

necessary.

3

AL. 11

ST. PRSS

SP outside limit

Static pressure exceeds

limit.

Continues to operate

and output.

AL. 12

CAP. TMP

CT outside limit

Capsule temperature is

outside range (–50 to

130°C).

Use heat insulation or

make lagging to keep

temperature within

range.

AL. 13

AMP. TMP

AT outside limit

Amplifier temperature

is outside range (–50 to

95°C).

AL. 30

RANGE

P over range

Output is outside upper

or lower range limit

value.

Outputs AO upper limit

or AO lower limit.

Check input and range

setting, and change

them as needed.

4

AL. 31

SP. RNG

SP over range

Static pressure exceeds

specified range.

Continues to operate

and output.

AL. 35

P. HI

P high alarm

Input pressure exceeds

specified threshold.

Check input.

5

AL. 36

P. LO

P low alarm

AL. 37

SP. HI

SP high alarm

Input static pressure

exceeds specified

threshold.

AL. 38

SP. LO

SP low alarm

AL. 39

TMP. HI

CT high alarm

Detected temperature

exceeds specified

threshold.

Check capsule

temperature.

AL. 40

TMP. LO

CT low alarm

AL. 50

P. LRV

Illegal P LRV

Specified value is

outside of setting range.

Holds output

immediately before

error occurred.

Check settings and

change them as

needed.

6

AL. 51

P. URV

Illegal P URV

AL. 52

P. SPN

Illegal P SPAN

AL. 53

P. ADJ

P SPAN trim err

Continues to operate

and output.

Adjust settings and

change them as

needed.

P ZERO trim err

AL. 54

SP. RNG

Illegal SP LRV

Continues to operate

and output holding

static pressure in %.

Check settings and

change them as

needed.

Illegal SP URV

Illegal SP SPAN