14 signal characterizer, Signal characterizer -26 – Yokogawa EJX440A User Manual

Page 36

Advertising
background image

<3. Operation>

3-26

IM 01C25T01-01E

3.5.14 Signal Characterizer

This function is used to compensate the output

for non-linear applications. The characterized

values are applied to the 4-20 mA output. For

the measured pressure, a maximum of nine

coordinates can be specified between 0-100%.

Perform the coordinate settings while the

S.C. at

S.C. menu parameter is Disabled.
To apply the settings to the output, set the

S.C.

parameter to

Enabled.

Note that the EJX rejects the activation of the

function by AL. 60 with the following transmitter’s

status:

• When the specified coordinates of x and y are

not incremental as the input increases.

• When the output mode of the output signal is

set as

Sq root; at the same time, the low cut

mode is set to

Linear.

F0343.ai

Y

X

100%

0%

100%

INPUT

OUTPUT

Input pressure in %

Characterized value

(ENTER)

EJX:

Num of points

0

5

HELP

DEL

ESC

ENTER

Enter 5 and press

ENTER (F4).

1. Device setup
4. Detailed setup
2. Signal condition
S.C. menu
2. Num of points

Example: Set the number of coordinates on the line graph

to 5.

1

‘ 5 ’

F0344.ai

EJX:

S.C. menu

1 S.C. Disabled

2 Num of points 5

3 point setting

4 X Start 0.00 %

5 Y Start 0.00 %

HELP

SEND

HOME ENTER

Press

SEND (F2) to send the

data to the transmitter, then

check to confirm that

disappears.

(SEND)

SEND

2

F0345.ai

1. Device setup
4. Detailed setup
2. Signal condition
S.C. menu
3. Point setting

Example: Set the first coordinates (X1, Y1) as (12, 14) in %.

(ENTER)

EJX:

Select one

1 List

2 Edit all points

3 Edit one point

4 Exit

ABORT ENTER

Select

Edit one point, and press

ENTER (F4).

1

(ENTER)

EJX:

Specify S.C. point to

set value

1

1

DEL

ABORT ENTER

Press

ENTER (F4).

2

(ENTER)

EJX:

X1

10.00 %

12

HELP

DEL

ESC

ENTER

Enter 12 and press

ENTER (F4).

3

‘ 12 ’

(ENTER)

EJX:

Y1

10.00 %

14

HELP

DEL

ESC

ENTER

Enter 14 and press

ENTER (F4).

4

‘ 14 ’

×2

F0346.ai

(ENTER)

EJX:

S.C

Disabled

Disabled

Enabled

HELP

DEL

ESC

ENTER

Select

Enabled, and press

ENTER (F4).

1. Device setup
4. Detailed setup
2. Signal condition
S.C. menu
1. S.C.

Example: Set the signal characterizer to

Enabled.

EJX:

S.C. menu

1 S.C. Enabled

2 Num of points 5

3 Point setting

4 X Start 0.00 %

5 Y Start 0.00 %

HELP

SEND

HOME ENTER

Press

SEND (F2) to send the data

to the transmitter, then check to

confirm that disappears.

(SEND)