1 blockage detection, 1 blockage detection -4 – Yokogawa EJX440A User Manual

Page 44

Advertising
background image

<4. Diagnostics>

4-4

IM 01C25T01-01E

4.2.2.1 Blockage Detection

Limit parameter

When the parameter based on pressure fluctuation

exceeds the preset value, EJX diagnoses an

impulse line as blockage and gives an alarm. The

threshold values are set to Limit parameter shown

in below table.

Table 4.2.2 Limit Parameter

#

Parameter

Threshold value

[1] Lim fDPmax

Threshold to detect “A

Blocking” by using

Ratio fDP

[2] Lim fDPmin

Threshold to detect “B

Blocking” by using

Ratio fDP

[3] Lim fSPlmax

Threshold to detect “Large

Fluct L” by using

Ratio fSPl

[4] Lim fSPlmin

Threshold to detect “L Side

Blocking” by using

Ratio fSPl

[5] Lim fSPhmax

Threshold to detect “Large

Fluct H” by using

Ratio fSPh

[6] Lim fSPhmin

Threshold to detect “H Side

Blocking” by using

Ratio fSPh

[7] Lim BlkFmax

Threshold to detect “H Side

Blocking” by using

BlkF

[8] Lim BlkFmin

Threshold to detect “L Side

Blocking” by using

BlkF

[9] Lim DPAvgmax Threshold to detect “ILDB over

range” by using

DPAvg and to

detect “Invalid Ref xx” by using

Ref DPAvg

[10] Lim DPAvgmin

Threshold to detect “ILDB over

range” by using

DPAvg and to

detect “Invalid Ref xx” by using

Ref DPAvg

Table 4.2.3 shows the default values at the factory

setting, which are different according to the model.

NOTE

• When ILBD is performed for the first time,

use the default value. If the pressure

fluctuation amplitude is low or a false alarm

is often generated after ILBD is performed,

change the values of Limit parameters

according to the procedure described in

4.2.2.10.

• Use Device description (DD) file for

parameter setting.

Table 4.2.3 Default Values of Limit Parameter

#

Parameter

Differential pressure

transmitter

Flange

mounted

differential

pressure

transmitter

Diaphragm

sealed differential

pressure/

pressure

transmitter

Absolute

pressure

transmitter

Gauge

pressure

transmitter

Gauge/

Absol ute

pressure

transmitter

EJX110A

EJX115A

EJX130A

EJX120A

EJX210A

EJX118A

Note 1

EJX438A

EJX310A

EJX430A

EJX440A

EJX510A

EJX530A

[1]

Lim fDPmax

3

3

10000

10000

10000

10000

10000

[2]

Lim fDPmin

0.3

0.3

0.3

0.3

0.3

0.3

0.3

[3]

Lim fSPlmax

5

10000

10000

10000

10000

10000

10000

[4]

Lim fSPlmin

0.5

0

0

0

0

0

0

[5]

Lim fSPhmax

5

10000

10000

10000

10000

10000

10000

[6]

Lim fSPhmin

0.5

0

0

0

0

0

0

[7]

Lim BlkFmax

0.6

10

10

10

10

10

10

[8]

Lim BlkFmin

-0.6

-10

-10

-10

-10

-10

-10

[9] Lim DPAvgmax

Note 2

1

1

1

1

1

1

1

[10] Lim DPAvgmin

Note 2

0.05

0.2

-1

-1

0.05

0.05

0.05

Note 1: The default values are set for level measurement. If EJX118A is applied to flow measurement, enter the same value to Limit

parameter [1] to [10] as those of EJX110A.

Note 2: It indicates the threshold value for “ILBD over range” (refer to 4.2.2.5).