<4. Diagnostics>

4-21

IM 01C25T01-01E

Table 4.3.2 HART Communicator Error Messages

Error message

Probable cause

Countermeasure

Invalid selection

Change the setting.

Value was too high

Set value is too high.

Value was too low

Set value is too low.

Too few data bytes received

In write protect mode

Operation is set in the Write Protect mode. —

Lower range value too high

LRV set point is too high.

Change the range.

Lower range value too low

LRV set point is too low.

Upper range value too high

URV set point is too high.

Upper range value too low

URV set point is too low.

Span too small

Set span is too small.

Applied process too high

Applied pressure is too high.

Adjust the applied pressure.

Applied process too low

Applied pressure is too low.

New lower range value pushed upper

range value over upper senser limit

The shift of URV according to the new LRV

setting exceeds USL.

Change the URV setting within

the range of USL.

Excess correction attempted

Amount of correction is too much.

Adjust the amount.

Lowercase conversion not succeeded

Characters are not convertible. e.g. %

Correct the setting.

Not in fixed current mode

The fixed current mode is desired but not

set in that mode.

Set in the fixed current mode.

In multidrop mode

Operation is set in the multi-drop mode.

Not write protect mode

Operation is set without a password.

Lower range value and upper range value

out of limits

Set value is outside lower or upper range

limit.

Change the setting.