IM 11M13A01-04E

8-15

8. Detailed Data Setting

8.7.2

Setting Periods over which Average Values are Calculated and Periods over which
Maximum and Minimum Values Are Monitored

The equipment enables the display of oxygen concentration average values and maxi-
mum and minimum values under measurement (see Section 10.1.1, later in this manual).
The following section describes how to set the periods over which oxygen concentration
average values are calculated and maximum and minimum values are monitored.

8.7.2.1 Procedure

Use the parameter-code table below to set the average, maximum and minimum oxygen
concentration values. Periods over which average is calculated and periods over which
maximum and minimum values are monitored can be set, ranging from 1 to 255 hours.
If the set ranges are beyond the limits specified, an “ERR” will be displayed.

Table 8.15 Parameter Codes for Average, Maximum and Minimum Values

Set item

Parameter code Set range Units

Periods over which average

F11

1 to 255

Hours

values are calculated
Periods over which maximum

F12

1 to 255

Hours

and minimum values are
monitored

T8.15E.EPS

8.7.2.2 Default Value

When the analyzer is delivered, or if data are initialized, periods over which average
values are calculated are set to one hour, and periods over which maximum and mini-
mum values are monitored are set to 24 hours.