5 contact input wiring, 1 cable specifications, 2 wiring procedure – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 62

Advertising
background image

IM 11M13A01-04E

5-6

5.5

Contact Input Wiring

The converter can execute specified function when receiving contact signals.

To use these contact signals, proceed wiring as follows:

Converter

Terminal box

DI-1

DI-C

DI-2

Contact input 1

Contact input 2

F5.7E.EPS

Hazardous Area

Non-hazardous Area

Figure 5.14 Contact Input Wiring

5.5.1

Cable Specifications

Use a 2-core or 3-core cable for this wiring. Depending on the number of input(s),
determine which cable to use.

5.5.2

Wiring Procedure

(1) M4 screws are used for the terminal of the converter. Each cable should be equipped

with the corresponding crimp contact.

(2) The ON/OFF level of this contact input is identified by the resistance. Connect a

contact input that satisfies the descriptions in Table 5.2.

Table 5.2

Identification of Contact Input ON/OFF

T5.2E.EPS

Closed

Open

Resistance

200

or less

100 k

or more