Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 152

Advertising
background image

IM 11M13A01-04E

11-10

(7) To restore the electronics, reverse the above removal procedures.

When restoring the electronics, do not get leadwire jammed in any part of the unit.

(8) Place the electronics and the printed-circuit board on which the fuse is installed

properly; these are directly connected with connectors.

(9) Tighten the four screws in their positions.

(10) Replace and tighten the display cover properly. If the cover is not tightened

sufficiently, the infrared switches will not operate correctly.

᭿ Fuse rating

Check the rating of the fuse and that it satisfies the following :
Maximum rated voltage

: 250 V

Maximum rated current

: 3.15 amperes

Type

: Time-lag fuse

Standards

: UL-, CSA- and VDE-approved

Part number

: A1113EF