1 general specifications, 1 standard specifications – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 20

Advertising
background image

IM 11M13A01-04E

2-2

2.1

General Specifications

2.1.1

Standard Specifications

Measured Object: Oxygen concentration in combustion exhaust gas and mixed gas

(excluding inflammable gases. May not be applicable corrosive gas such
as ammonia and chlorine.)

The sampling gases containing a corrosive gas such as ammonia or
chlorine may be applicable to our oxygen gas analyzer. In this case,
contact with YOKOGAWA and its agency.

Measured System: Zirconia system

Oxygen Concentration: 0.01 to 100 vol% O

2

Measurement Range: Any setting in the range of 0 to 5 through 0 to 100 vol% O

2

(in 1 vol%O

2

), or partial range

Display Range: 0 to 100 vol% O

2

Warm-up Time: Approx. 20 min.

Repeatability:

Ϯ 0.5% Maximum value of range set . (at least 0 to 5 vol% O

2

, and less

than 0 to 25 vol% O

2

range)

Ϯ 1% Maximum value of range set. (at least 0 to 25 vol% O

2

and up to

0 to 100 vol% O

2

range)

Linearity: (Excluding standard gas tolerance)

(Use oxygen of known concentration (in the measuring range) as the zero and
span calibration gas.)

Ϯ 1% Maximum value of range set.; range at least 0 to 5vol% O

2

, and up to 0

to 25 vol% O

2

range

(Sample gas pressure: within 4.9 kPa)
Ϯ 3% Maximum value of range set.; at least 0 to 25 vol% O

2

or and less than

0 to 50 vol% O

2

range

(Sample gas pressure: within 0.49 kPa)
Ϯ 5% Maximum value of range set.; at least 0 to 50 vol %O

2

and up to 0 to

100 vol% O

2

range

(Sample gas pressure: within 0.49 kPa)

Drift: (Excluding the first two weeks in use)

both zero and span 2% Maximum value of range set/month

Response Time: Response of 90% within 5 seconds. (Measures after gas is introduced
from calibration gas inlet and analog output start changing.)