8. Detailed Data Setting

8.

Detailed Data Setting

8.1

Setting Display Item

Display items are defined as items displayed on the basic panel display.

Parameter code A00 or F08 is used to set the display items as shown in Table 8.1. The
oxygen concentration is set at the factory before shipment. In addition, if the data
initialization is performed, the oxygen concentration will be set.

Table 8.1 Display Item

T8.1E.EPS

Values set with

Items displayed on the basic panel display

A00 or F08
0

Indicates the oxygen concentration.

1 or 2

For humidity analyzers only. (if 1 or 2 is set for the oxygen analyzer,
"0.0" is only displayed on the basic panel display.)

3

Displays an item for the current output.
If the output damping has been set for the current output,
values involving the output damping are displayed.

CAUTION

If you set “3” in the parameter code A00 or F08, be sure to select “Oxygen Concentra-
tion” in the following mA output setting (see Section 8.2, “Current Output Setting”).