3. Installation

3.

Installation

This chapter describes installation of the following equipment:

3.1

Model ZR202S Zirconia Oxygen Analyzer

3.2

Model ZA8F Flow Setting Unit

3.3

Case Assembly for Calibration-gas Cylinder (E7044KF)

3.1

Installation of the Zirconia Oxygen Analyzer

3.1.1

Area

The following should be taken into consideration when installing the probe:

(1) Easy and safe access to the probe for checking and maintenance work.
(2) Ambient temperature of not more than 55˚C, and the terminal box should not be

affected by radiant heat.

(3) A clean environment without any corrosive gases.
(4) No vibration.
(5) The measurement gas satisfies the specifications described in Chapter 2.
(6) No measured gas-pressure fluctuations.

CAUTION

• The ambient temperature of ZR202S Integrated type Explosion Zirconia Oxygen

Analyzer should be between

؊20˚C and 55˚C.