Alarm-related items in group d – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 133

Advertising
background image

IM 11M13A01-04E

10-15

10. Other Functions

Alarm-related Items in Group D

T10.6D.EPS

Code Item

Tuning

Engineering unit Default

setting

D01 Oxygen concentration, high-high

0 to 100

% O

2

100% O

2

alarm setpoint

D02 Oxygen concentration, high-limit

0 to 100

% O

2

100% O

2

alarm setpoint

D03 Oxygen concentration, low-limit

0 to 100

% O

2

0% O

2

alarm setpoint

D04 Oxygen concentration, low-low

0 to 100

% O

2

0% O

2

alarm setpoint

D05

D06

D07

D08

D11

D12

D13

D14

D30 Oxygen concentration alarm

0 to 9.9

% O

2

0.1% O

2

hysteresis

D31

D32

D33 Delayed alarm action

0 to 255

Seconds 3 seconds

D41 Oxygen concentration,

0: Not detected

Not detected

high-high alarm detection

1: Detection

D42 Oxygen concentration, high-limit

0: Not detected

Not detected

alarm detection

1: Detection

D43 Oxygen concentration, low-limit

0: Not detected

Not detected

alarm detection

1: Detection

D44 Oxygen concentration, low-low

0: Not detected

Not detected

alarm detection

1: Detection

D45

D46

D47

D48

D51

D52

D53

D54