2 operational data initialization, Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 126

Advertising
background image

IM 11M13A01-04E

10-8

10.2 Operational Data Initialization

Individual set data initialization enables you to return to the default values set at the time
of delivery. There are two types of initializations: an all set-data initialization and a
parameter-code-based initialization. Table 10.4 lists the initialization items by a param-
eter code, and default values.

Table 10.4 Parameter Codes for Initialization

Parameter code

Data to be initialized

F30

All data

F31

Data in Group A

F32

Data in Group B

F33

Data in Group C

F34

Data in Group D

F35

Data in Group E

F36

Data in Group F

T10.4EPS

CAUTION

When Data in Group Fare initialized by the parameter code of F36, F01 and F02 and
F08 and F10 cannot be initialized.