For the safe use of this equipment

WARNING

Requirements for explosionproof use:

The ambient temperature is in the range -20 to +55

ЊC. The surface temperature of the

ZR202S is not over the temperature class T2 (300

ЊC)͓note 1͔

* The surface temperature of the amplifier box does not exceed 70

ЊC.

CAUTION

The cell (sensor) at the tip of the probe is made of ceramic (zirconia element). Do not
drop the equipment or subject it to pressure stress.

• Do NOT allow the sensor (probe tip) to make contact with anything when installing

the analyzer.

• Avoid water dropping directly on the probe (sensor) of the analyzer when installing it.
• Check the calibration gas piping before introducing the calibration gas to ensure that

there is no leakage of the gas. If there is any leakage of the gas, the moisture drawn
from the measured gas may damage the sensor.

• The probe (especially at the tip) becomes very hot. Be sure to handle it with gloves.

DANGER

EXAxtZR is very heavy. Be sure not to accidentally drop it. Handle safely to avoid

injury.

Connect the power supply cord only after confirming that the supply voltage matches
the rating of this equipment. In addition, confirm that the power is switched off when
connecting power supply.

Some process gas is dangerous to people. When removing this equipment from the
process line for maintenance or other reasons, protect yourself from potential poisoning
by using a protective mask or ventilating the area well