Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 80

Advertising
background image

IM 11M13A01-04E

7-16

7.9.1

Contact Output Check

Follow Table 7.10 to check the contact output. The table uses an example with contact
output 1.

Table 7.10 Checking Contact Output

Display after the password has been entered.

Touch the [ ] key to switch to Group F.

Touch the [>] key to move the position of the digit that
is flashing to the right one digit.

Touch the [ ] key to enter 1.

Touch the [

ENT

] key to have 0 flash. The contact is then open.

Touch the [ ] key to set 1 (one).

Touch the [

ENT

] key. The flashing continues.

Touch the [

ENT

] key again to stop the flashing, and the

contact will be closed.

Touch the [

ENT

] key once again to switch to the parameter code selection display.

The contact then returns to the original state.

Touch the [>] key together with the [

ENT

] key to return to the basic panel display.

(This is not required if you proceed to make another setting.)
(The displayed numeric characters indicate the measurement gas concentration.)

T7.9E.EPS

Switch operation

Display

Description

>

>

>

>

>

>

>

>

>

>

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

A

01

G

01

G

0

1

G

1

1

0

1

1

1

G

1

1

Basic
panel
display

The symbol [ ] indicates that the corresponding keys are being touched, and the light characters indicate flashing.

ٙ

ٙ

ٙ

CAUTION

• If you conduct an open-close check for the contact output 2, Error 1 (cell voltage

failure) or Error 2 (heater temperature abnormal) will occur. This is because the built-
in heater power of the detector, which is connected to contact output 2, is turned off
during the above check. So, if the above error occurs, reset the equipment or turn the
power off and then back on to restart (refer to Section 10.4, “Resetting” later in this
manual).