Caution, 3 selection of output mode, 4 default values – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 91

Advertising
background image

IM 11M13A01-04E

8-3

8. Detailed Data Setting

8.2.3

Selection of Output Mode

You can select a linear or logarithmic output mode. The former provides linear charac-
teristics between the analog output signal and oxygen concentration.

8.2.4

Default Values

When the analyzer is delivered or data are initialized, the output current settings are by
default as shown in Table 8.3.

Table 8.3 Output Current Default Values

Item

Default setting

Min. oxygen concentration

0% O

2

Max. oxygen concentration

25% O

2

Output damping constant

0 (seconds)

Output mode

Linear

T8.3E.EPS

CAUTION

• When you select logarithmic mode, the minimum output remains constant at 0.1% O

2

,

and the parameter C11 display remains unchanged.

Advertising