Sa-cal, Span, Zero – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 117: Calend, Caution, 2 calibration

Advertising
background image

IM 11M13A01-04E

9-9

9. Calibration

9.2.2

Calibration

9.2.2.1Manual Calibration

For manual calibration, consult Section 7.10, “Calibration” earlier in this manual.

9.2.2.2Semi-automatic Calibration

(1) Calibration startup using infrared switches

Table 9.3 Semi-automatic Calibration Procedure

Switch operation

Display

Brief description

>

ENT

b

11

>

ENT

SA-CAL

>

ENT

SPAn

/20.84

>

ENT

ZEro

/0.89

>

ENT

CALEnd

>

ENT

T9.3E.EPS

Change the parameter code to B11. (Previous operations omitted)

Touch the [

ENT

] key to display "

SA-CAL

" (Semi Auto CAL).

Touch the [

ENT

] key again to open the span-gas solenoid valve.

The span gas then flows. "

SPAN

" and the currently measured value are alternately

displayed. If the "

output hold

" is set, the output hold will start at this time.

If the set calibration time elapses, the span-gas solenoid valve closes automatically,
the zero-gas solenoid valve opens and the zero gas flows. "

ZERO

" and the currently

measured value are displayed alternately.

End If the set calibration time elapses, the zero-gas solenoid valve then closes
automatically. The ’’CALEND’’ flashes until the set output stabilization time elapses.

If the output stabilization time elapses, the basic panel display then appears.
Output holding will be released.

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Basic panel
display

The symbol ( ) indicates that the key is being touched. Light characters indicate that the digits are flashing.

/ indicates that both are displayed alternately.

(2) To start calibration using an input contact, follow these steps:

• Make sure that Calibration start has been selected in the Input contacts display (see

Section 8.4, earlier in this manual).

• Apply an input contact to start calibration.

(3) To stop calibration midway, follow these steps:

(1) Touch the [>] key and [ENT] key together. The calibration will stop and the

output stabilization time will be set up.

(4) Touch the [>] key once again to return to the basic panel display and the analyzer

will be in normal measurement.

9.2.2.3Automatic Calibration

No execution operations are required for automatic calibration. Automatic calibration
starts in accordance with a preset start date and time. Calibration is then executed at
preset intervals.

CAUTION

Before conducting a semi-automatic or automatic calibration, run the automatic calibra-
tion unit beforehand to obtain a calibration flow of 600

Ϯ 60 ml/min.