Wiring, 1 general, Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 57

Advertising
background image

IM 11M13A01-04E

5-1

5. Wiring

5.

Wiring

This chapter describes wiring procedures necessary for the EXAxtZR Integrated-type
Explosionproof Zirconia Oxygen Analyzer.

5.1

General

CAUTION

Be sure to read Sections 3.1.2 to 3.1.4 where the important information on wiring is
provided.

CAUTION

• Never supply current to the equipment or any other device constituting a power circuit

in combination with the equipment, until all wiring is completed.

• This product complies with CE marking.

Where a performance suit for CE marking is necessary, the following wiring proce-
dure is necessary.

1. Install an external switch or circuit breaker to the power supply of the equipment.
2. Use an external switch or circuit breaker rated 5A and conforms to IEC 947-1 or IEC

947-3.

3. It is recommended that the external switch or circuit breaker be mounted in the same

room as the equipment.

4. The external switch or circuit breaker should be installed within the reach of the

operator, and marked as the power supply switch of this equipment.

Wiring procedure

Wiring should be made according to the following procedure:
1. Be sure to connect the shield of the shielded line to FG terminal of the analyzer.
2. The most outer sheath of the signal line and the power cable should be stripped off

to the minimum necessary length.

3. Signal will be affected by noise emission when the signal lines, power cable are

located in the same conduit. When using a conduit, signal lines should be installed in
the separate conduit from power cables. Be sure to ground the metal conduit.

4. Mount the attached two blind plugs to unused cable connection gland(s) of the

equipment.

5. The cables indicated in Table 5.1 are used for wiring.
6. After completing the wiring, screw the cover in the terminal box body and secure it

with a lock screw.

Table 5.1 Cable Specifications

Terminal name of equipment

Name

Needs for shields Cable type

Number or wires

L, N,

Power supply

CVV

2 or 3 *

AO+, AO-

Analog output

CVVS

2

DO-1, DO-2

Contact output

CVV

2 to 8

DI-1, DI-2, DI-C

Contact input

CVV

3

Note *: When the case is used for protective grounding, use a 2-wire cable.

T5.1E.EPS

Note

• Select an appropriate cable O.D. for the cable gland size.
• Protective grounding should have the grounding resistance of 100

⍀ or less.