1.1

< EXAxtZR > System Configuration

The system configuration should be determined by the conditions; e.g. whether the
calibration is to be automated, and whether flammable gas is present and requires safety
precautions. The system configuration can be classified into two basic patterns as
follows:

1.1.1

System 1

This system is for monitoring and controlling oxygen concentration in the combustion
gases of a large-size boiler or heating furnace. Instrument air (clean and dry air of
oxygen concentration 21%) is used as the reference gas and the span gas for calibration.
Zero gas is fed from a cylinder during calibration. The gas flow is controlled by the
ZA8F flow setting unit (for manual valve operation).

~

ZR202S Integrated type Explosionproof
Zirconia Oxygen Analyzer

F06.EPS

ZA8F flow setting unit

Reference gas

Calibration gas

Needle

valve

Flowmeter

Instrument air

Air Set

Pressure regulator

Zero gas

cylinder

Stop valve

or

Check valve

Span gas (*4)

Non-hazardous area

Hazardous area

100 to 240 V AC