2 calibration gas – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 110

Advertising
background image

IM 11M13A01-04E

9-2

0.1

0.5

1

5

10

21.0

50

100

120

100

80

60

40

20

0

-20

-40

0.51% 0

2

,81.92mV(Zero origin of calibration)

21.0% O

2

, 0mV

(Span origin of calibration)

Oxygen concentration (vol %)

Cell
voltage
(mV)

F9.1E.EPS

Figure 9.1 Oxygen Concentration in a Measurement Gas vs. Cell Voltage (21 %

Equivalent)

The measurement principles of a zirconia oxygen analyzer have been described above.
However, the relationship between oxygen concentration and the electromotive force of
a cell is only theoretical. Usually, in practice, a sensor shows a slight deviation from the
theoretical value. This is the reason why calibration is necessary. To meet this require-
ment, an analyzer calibration is conducted so that a calibration curve is obtained, which
corrects the deviation from the theoretical cell electromotive force.

9.1.2

Calibration Gas

A gas with a known oxygen concentration is used for calibration. Normal calibration is
performed using two different gases: a zero gas of low oxygen concentration and a span
gas of high oxygen concentration. In some cases, only one of the gases needs to be used
for calibration. However, even if only one of the gases is normally used, calibration
using both gases should be done at least once.

The zero gas normally used has an oxygen concentration of 0.95 to 1.0 vol% O

2

with a

balance of nitrogen gas (N

2

). The span gas widely used is clean air (at a dew-point

temperature below -20˚C and free of oily mist or dust, as in instrument air).