Calibration-related items in group b – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 131

Advertising
background image

IM 11M13A01-04E

10-13

10. Other Functions

Calibration-related Items in Group B

Code

Item

Tuning

Engineering unit Default setting

B01

Zero-gas concentration

0.3 to 100

% O

2

1% O

2

B02

Span-gas concentration

4.5 to 100

% O

2

21% O

2

B03

Calibration mode

0: Manual calibration

Manual calibration

1: Semi-automatic and manual

calibration

2: Automatic, semi-automatic, and

manual calibration

B04

Output stabilization time

0 minutes, 0 seconds to 60 minutes, 59

MM.SS

10 minutes, 0 seconds

seconds

B05

Calibration time

0 minutes, 0 seconds to 60 minutes, 59

MM.SS

10 minutes, 0 seconds

seconds

B06

Calibration period

0 days 0 hours to 255 days 23 hours

DD.HH

30 days, 0 hours

B07

Start time

YY.MM.DD/HH.MM 00.01.01.00.00

B08

Calibration

0: Zero and span

Zero and span

1: Span only

2: Zero only

B09

Calibration concentration

Display only

% O

2

measurement

B10

Manual calibration

B11

Semi-automatic calibration

T10.6B.EPS