Warning – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 153

Advertising
background image

IM 11M13A01-04E

11-11

11.

Inspection and Maintenance

11.3 Replacement of Flowmeter for Autocalibration Unit

(1) Remove pipe holding piping fitting
(2) Remove bolts holding flowmeter, and replace it. A white back plate (to make the

float easy to see) is attached. The end of the pin holding down the back plate must
be on the bracket side.

(3) Replace piping, and fix M6 bolts between brackets. *1

*1 : When disassembling and reassembling, mark original positions, and tighten an

extra 5-10˚ when reassembling. After tightening, do a liquid leakage test.

Vertical mounting

Horizontal mounting

C

C

B

B

C'

B'

A'

A

A

Zr20h_g0.eps

Connect piping pairs A-A

Ј

, B-B

Ј

, C-C

Ј

, D-D

Ј

Fixing screw pairs

Flame Arrester

Fitting

Flame Arrester

Fitting

Figure 11.10 Fixing Flowmeter

WARNING

Do not loosen or remove any Flame Arrester of gas inlet/outlet during piping. The

detector modification or parts replacement by other than authorized representative of

Yokogawa Electric Corporation is prohibited and will void KEMA Flameproof

Certification, Factory Mutual Explosionproof Approval and Canadian Standards

Explosionproof Certification.