3.1.4 CSA Explosionproof Type

ZR202S–C Detector for use in hazardous area:

Note 1: Explosionproof for Class I, Division 1, Groups B, C and D

Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G

Enclosure Rating: TYPE 4X

Temperature class: T2

Note 2: Wiring

• All wiring shall comply with Canadian Electrical Code Part 1 and Local Electrical
Code.

• In hazardous area, wiring shall be in conduits as shown in the figure.

WARNING: SEAL ALL CONDUITS WITHIN 50 cm OF THE ENCLOSURE.

UN SELLE DOIT ÊTRE INSTALLÊ Á MOINS DE 50 cm DU
BîTIER.

Hazardous Area Division 1

50cm MAX.

Non-hazardous
Area

Non-hazardous
Area

Conduit

Sealimg Fitting

50cm MAX.

Figure 3.2 ZR202S DETECTOR

Note 3: Operation

• Keep the “WARNING” label to the Detector.

WARNING: OPEN CIRCUIT BEFORE REMOVING COVER. REFER TO

IM 11M13A01–04E. USE AT LEAST 80˚C HEAT RESISTANT
CABLES.

OUVRIR LE CIRCUIT AVANT D’fENLEVER LE COUVERCLE. UTILISEZ
DES CÁBLES RÊSISTANTS Á 80˚C MINIMUM. VEUILLEZ VOUS
RÊFÊRER AU IM 11M13A01–04E.

• Take care not to generate mechanical sparking when accessing to the detector and
peripheral devices in hazardous area.

Note 4: Maintenance and Repair

• The detector modification or parts replacement by other than authorized
representative of Yokogawa Electric Corporation is prohibited and will void Canadian
Standards Explosionproof Certification.