2. Specifications

2.4

Other Equipment

2.4.1

Stop Valve (part no. L9852CB or G7016XH)

This valve is mounted on the calibration gas line in the system to allow for one-touch

calibration. This is applied to a system configuration (System 1).

• Standard Specification

Connection: Rc 1/4 or 1/4 NPT(F)

Material: SUS 316 (JIS)

Weight: Approx. 80 g

Joint: R

C

1/4, Material: SUS316 (JIS)

Joint: 1/4 NPT, Material: SUS316 (JIS)

T14.EPS

Stop valve

Nipple

Description

Part No.

L9852CB

G7016XH

R

1/4, Material: SUS316 (JIS)

1/4 NPT, Material: SUS316 (JIS)

Description

Part No.

G7209XA

K9470ZN

Nipple.EPS

40

Rc1/4 or 1/4NPT

55

(Full open length)

43

G7209XA 2-Rc1/4
K9470ZN 2-1/4NPT

L9852CB 2-Rc1/4
G7016XH 2-1/4NPT

Analyzer

Approx. 100

2.4.2

Check Valve (part no. K9292DN or K9292DS)

This valve is mounted on the calibration gas line (directly connected to the detector).

This is applied to a system based on the system configuration (System 2 and 3).

This valve prevents the process gas from entering the calibration gas line. Although it
functions as the stop valve, operation is easier as it does not require opening/closing at
each calibration.

Screw the check valve into the calibration gas inlet of the detector instead of the stop
valve.

• Standard Specifications

Connection: Rc 1/4 or 1/4 NPT(F)

Material: SUS304 (JIS)

Pressure: Between 70 kPa G or more 350 kPa G or less

Weight: Approx. 40 g

Joint: RC 1/4, Material: SUS304 (JIS)

Joint: 1/4 NPT, Material: SUS304 (JIS)

T2.10E.EPS

Part No.

K9292DN

K9292DS

Description