IM 11M13A01-04E

8-6

8.3.2

Preference Order of Output Hold Value

The output hold value takes the following preference order:

During error occurrence

During calibration

During maintenance

During warm-up

Preference order (high)

8.2.2E.siki

For example, if the output current is set to 4 mA during maintenance, and no output-
hold output during calibration is preset, the output is held at 4 mA in the maintenance
display. However, the output hold is released at the time of starting the calibration, and
the output will be held again at 4 mA after completing the calibration and when the
output stabilization time elapses.

8.3.3

Output Hold Setting

Table 8.5 lists parameter codes with set values for individual set items.

Table 8.5 Parameter Codes for Output Holding

Set items

Parameter code

Set value

During warm-up

C04

0: 4 mA

1: 20 mA

2: Holds set values.

During maintenance C05

0: Without hold feature

1: Last measured value.

2: Holds set values.

During calibration

C06

0: Without hold feature

1: Last measured value.

2: Holds set values.

During error occurrence C07

0: Without hold feature

1: Last measured value.

2: Holds set values.

T8.5E.EPS

8.3.4

Default Values

When the analyzer is delivered, or if data are initialized, output holding is by default as
shown in Table 8.6.

Table 8.6 Output Hold Default Values

Status

Output hold

During warm-up

4 mA

During maintenance

Last measured value.

During calibration

Last measured value.

Error occurrence

Holds output at a preset value.

The default preset value is 3.4 mA only for output hold on error occurrence and others, it is 4 mA.

T8.6E.EPS