3.1.3 FM Explosionproof Type

ZR202S–B Analyzer for use in hazardous area:

Note 1: Explosionproof for Class I, Division 1, Groups B, C and D

Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G

Enclosure Rating: NEMA 4X

Temperature class: T2

Note 2: Wiring

• All wiring shall comply with National Electrical Code ANSI/NEPA 70 and Local
Electrical Code.

• In hazardous area, wiring shall be in conduits as shown in the figure.

WARNING: SEAL ALL CONDUITS WITHIN 18 INCHES OF THE

ENCLOSURE.

Hazardous Area Division 1

18 inches (475mm) MAX.

18 inches (475mm) MAX.

Non-hazardous
Area

Non-hazardous
Area

Conduit

Sealimg Fitting

Figure 3.1 ZR202S Detector

Note 3: Operation

• Keep the “WARNING” label to the Detector.

WARNING: OPEN CIRCUIT BEFORE REMOVING COVER. INSTALL IN

ACCORDANCE WITH THE INSTRUCTION MANUAL

IM 11M13A01–04E. USE AT LEAST 80˚C HEAT
RESISTANT CABLES.

• Take care not to generate mechanical sparking when accessing to the detector and
peripheral devices in hazardous area.

Note 4: Maintenance and Repair

• The detector modification or parts replacement by other than authorized
representative of Yokogawa Electric Corporation is prohibited and will void Factory
Mutual Explosionproof Approval.