IM 11M13A01-04E

8-5

8. Detailed Data Setting

F8.1E.EPS

Switch operations

Display

Output hold time
during calibration

>

ENT

b

1

0

>

ENT

CAL

>

ENT

CAL

>

ENT

SPAn

Y

>

ENT

21.00

%

>

ENT

OPEn/20.84

>

ENT

20.84

%

>

ENT

ZEro

Y

>

ENT

0.98

%

>

ENT

OPEn/0.89

>

ENT

0.89

%

>

ENT

CALEnd

>

ENT

b

10

>

ENT

Measured-value display

>

Figure 8.1 Definition During Calibration

In a semi-automatic calibration, “during calibration” is the time, starting when a calibra-
tion instruction is executed with an infrared switch or a contact input, to make a series
of calibrations, until the preset output stabilization time elapses.

In an automatic calibration, “during calibration” is the time, starting when automatic
calibration is carried out at the calibration start time, until the preset output stabilization
time elapses.

(4) “Error” appears when Error 1 to Error 4 are being issued