7.9.3

Checking Input Contacts

Follow Table 7.12 to check the input contacts. The table uses an example with input
contact 1.

Table 7.12 Checking Input Contacts

T7.11E.EPS

>

>

>

>

>

>

Ͼ

Ͼ

ϾϾ

Ͼ

Ͼ

ENT

ENT

ENT

ENT

ENT

ENT

Switch operation

Display

Description

A

01

G

01

G

0

1

G

2

1

0

G

21

The symbol [ ] indicates that the corresponding keys are being touched, and the light characters indicate flashing.

Display after the password has been entered.

Touch the [ ] key to switch to Group G.

Touch the [>] key to move the position of the digit
that is flashing to the right one digit.

Touch the [ ] key to enter 2.

Touch the [

ENT

] key. 0 is displayed with the contact open.

If the contact is closed, the display will be 1 (one).
This enables you to check whether or not the wiring
connections have been properly made or not.

Touch the [>] key together with the [

ENT

] key to return to

the basic panel display.