IM 11M13A01-04E

8-10

8.5

Output Contact Setup

8.5.1

Output Contact

Mechanical relays provide contact outputs. Be sure to observe relay contact ratings. (For
details, consult the specifications). The following sets forth the operation mode of each
contact output. Output contact 1 you can select open or closed contact when the contact
is "operated.". For output contact 2, output contact 2 is closed. The relay for output
contact 1 is energized when its contacts are closed and vice versa. Accordingly, when no
power is supplied to the equipment, those contacts remain open. In addition, the relay
for output contact 2 is energized when the corresponding contact is open and de-
energized when that contact is closed.

Table 8.9 Setting Output Contacts

Operating state

When no power
is applied to this
equipment

Output contact 1

Open (deenergized) or closed

Open

(energized) selectable.

Output contact 2

Closed (deenergized) only.

Closed

T8.9E.EPS