7.7

Output Range Setting

This section sets forth analog output range settings. For details, consult Section
8.2,”Current Output Settings,” later in this manual.

7.7.1

Minimum Current (4 mA) and Maximum Current (20 mA) Settings

Use the parameter codes C11 to set the oxygen concentration at 4 mA and C12 to set the
oxygen concentration at 20 mA. The following shows where 10% O

2

is set at 4 mA and

20% O

2

at 20 mA.

Table 7.7 Minimum and Maximum Value Setting Procedure