8 checking current loop – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 78

Advertising
background image

IM 11M13A01-04E

7-14

7.8

Checking Current Loop

The set current can be output as an analog output. This enables the checking of wiring
between the converter and the receiving instrument. Current loop checking is performed
using parameter code G01.

Table 7.8 Checking Current Loop

T7.7E.EPS

Switch operation

Display

Description

>

>

>

>

>

>

>

>

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

A

01

G

01

0

0.0

1

0.0

1

0.0

1

0.0

G

01

Basic
panel
display

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Display after the password has been entered.

Touch the [ ] key to switch to Group F.

Touch the [

ENT

] key. The output current remains preset

with the output-hold feature (Section 2.3).

Touch the [ ] key to set the numeric value 1 (to set a 10-mA output).

Touch the [

ENT

] key to have all the digits flash.

Touch the [

ENT

] key again to stop the flashing.

A 10-mA output is then issued.

Touch the [

ENT

] key once again to switch

to the parameter code selection display.

At that point, the output current returns to the normal value.

Touch the [>] key together with the [

ENT

] key

to return to the basic panel display.

The symbol [ ] indicates that the corresponding keys are being touched, and the light characters indicate flashing.

ٙ

ٙ