IM 11M13A01-04E

3-7

3. Installation

3.1.7

Installation of the Probe

CAUTION

• The cell (sensor) at the tip of the detector is made of ceramic (zirconia). Do not drop

the detector, as impact will damage it.

• A gasket should be used between the flanges to prevent gas leakage. The gasket

material should be heatproof and corrosion-proof, suited to the characteristics of the
measured gas.

The following should be taken into consideration when mounting the general-use
detector:

<General-use detector>
(1) Make sure that the cell mounting screws (four) at the probe tip are not loose.
(2) Where the detector is mounted horizontally, the calibration gas inlet and the refer-

ence gas inlet should face downward.

3.1.8

Installation of the Probe Protector (ZO21R)

<Detector with a probe protector (Model ZO21R-L-

ᮀᮀᮀ-ᮀ *B for dust wear protect>

The detector is used with a probe protector to prevent the sensor from being worn by
dust particles when there is a high concentration of dust and gas flow exceeds 10m/s
(fine-carbon boiler or fluid-bed furnace).

(1) Put the gasket that is provided by user between the flanges, and mount the probe

protector in the probe insertion hole. The probe protector should be installed so that
the notch comes to the downstream of the measurement gas flow.

(2) Make sure that the cell assembly mounting screws (four) at the probe tip are not

loose.

(3) Where the detector is mounted horizontally, the calibration/reference gas inlet should

face downward.

Direction of the measured gas flow

Mount the protector so that the notch
is on the downstream side of gas flow.

Unit: mm

F3.4E.EPS

1050

Gasket (t1.5)

Calibration gas inlet
Reference gas inlet

Detector top

Figure 3.4 Mounting of detector with a probe protector (Dust wear protect)